6.5

Incubator operation

This chapter describes how to set which modes and settings in incubator operation.
6.5.1
Overview of incubator operation
When the hood is closed, the device is automatically operated as an incubator. The
device displays the user interface for incubator operation.
Fig. 35
No.
1
2
3
6.5.1.1
Condensation management
High humidity settings may cause the inside of the hood to fog up. To limit fogging,
the ClearView function can be activated. If the ClearView on button is activated in
the therapy bar, the radiant warmer operates at low power. This warms the hood,
reducing fogging.
Instructions for use Babyleo TN500 SW 1.0n
3
Standard view in incubator operation with air temperature mode activated
Designation
Air temp.
ClearView on
Therapy bar
2
Description
Shows the current air temperature measure-
ment in the patient compartment.
Reduces condensation on the inside of the
hood.
Shows the available modes in incubator oper-
ation. In this example, air temperature mode is
activated.