The radiant warmer starts up with 100 % power. Within a period of 14 minutes and
30 seconds the power is gradually reduced. In pre-warm mode the therapy control
displays either Pre or a value, depending on the radiant warmer power.
Fig. 37
No.
1
Instructions for use Babyleo TN500 SW 1.0n
1
Pre-warm mode
Designation
Pre
Description
Shows that pre-warm mode is activated.
Operation
73