10.7

Removing liquid from the heating plate

If large quantities of liquids are spilled in the device, the liquid must be removed
immediately. The liquid must be removed before the device is disassembled and
moved. If necessary, specialized service personnel must check if the liquid has
penetrated into the device.
Fig. 81
WARNING
Risk due to penetrating liquid
If too much liquid is on the heating plate, the liquid may penetrate into the device
and cause damage to the device. If the device fails, the patient is put at risk.
► Make sure that no liquid penetrates into the device.
► If the liquid level on the heating plate reaches the upper edge of the fan impeller
receptacle, stop using the device and contact the specialized service personnel.
Procedure:
1 Switch off the device on the display unit using the Start/standby key .
2 Switch off the device with the main switch on the rear of the main column.
3 Disconnect the mains plug.
4 Remove any liquid on the mattress and the mattress tray.
5 Remove the mattress and mattress tray.
6 Remove any liquid on the bed support.
7 Remove the bed support and the T-bars.
8 Remove any liquid from the warm air duct covers.
9 Remove the warm air duct covers.
10 Check if the fan impeller is in the liquid and remove the fan impeller.
11 Check if there is liquid on the heating plate. The heating plate may be hot.
12 If the liquid level reaches the upper edge of the fan impeller receptacle (1),
contact the specialized service personnel. The specialized service personnel
must check if liquid has penetrated into the device.
164
Heating plate
Instructions for use Babyleo TN500 SW 1.0n