Skin Temp. Sensor Faulty, Heating; Skin Temperature Difference High; Skin Temperature Difference Low; Skin Temperature High - Dräger Babyleo TN500 Instructions For Use Manual

Hide thumbs Also See for Babyleo TN500:
Table of Contents

Advertisement

Alarm pri-
Alarm
ority
!
217
Prewarming in progress
!
215
Rotary knob faulty
!
117
Scale calibration required
!
115
Scale communication error A scale communication error
!
116
Scale disconnected
!
114
Scale faulty
!
211
Secondary speaker faulty
!! 100
Skin mode not confirmed
!! 117
Skin temp. sensor faulty,
heating off
!! 19
Skin temperature differ-
ence high
!! 18
Skin temperature differ-
ence low
!! 213

Skin temperature high

Instructions for use Babyleo TN500 SW 1.0n
Cause
The device is warming up.
The rotary knob has been
pressed for more than 10
seconds.
An internal fault occurred.
occurred.
An internal fault occurred.
The scale cable is not con-
nected.
An internal fault occurred.
An internal fault of the sec-
ondary acoustic alarm sys-
tem occurred.
The mode was selected, but
it was not confirmed.
At least one of the skin tem-
perature sensors is faulty.
The current temperature dif-
ference between central skin
temperature (yellow) and
peripheral skin temperature
(white) is above the set
upper alarm limit.
The current temperature dif-
ference between central skin
temperature (yellow) and
peripheral skin temperature
(white) is below the set lower
alarm limit.
The tolerated deviation for
central skin temperature (yel-
low) has been exceeded.
The current central skin tem-
perature is too high.
Troubleshooting
Remedy
Wait until prewarming is fin-
ished.
To deactivate prewarming,
switch to manual mode.
Release the rotary knob.
Contact specialized service
personnel.
Disconnect the scale cable.
Contact specialized service
personnel.
Make sure that the scale
cable is correctly connected.
Disconnect the scale cable.
Contact specialized service
personnel.
Make sure that the scale
cable is correctly connected.
Disconnect the scale cable.
Contact specialized service
personnel.
Contact specialized service
personnel.
Select skin mode and press
the rotary knob to confirm.
Replace the faulty skin tem-
perature sensor.
Contact specialized service
personnel.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
159