8.1.2
Recommended air temperature
If patient data, such as date of birth, gestational age and weight, is entered, then
the device uses this data to determine an air temperature range for the patient.
The suggested air temperature range is based on a naked patient and on the
following settings:
– ClearView on
– Humidity with the setting Auto
Procedure:
1 Enter the date of birth.
2 Enter the gestational age in weeks and days.
3 Select the actual weight range from the picklist.
Example: The current weight of the patient is 1057 g (2.33 lbs). The range
1000–1100 g (2.2–2.43 lbs) is selected from the pick list.
 The recommended air temperature range (1) for the patient is displayed.
Fig. 75
4 Close the System setup dialog window.
5 Touch Air mode in the therapy bar and confirm with the rotary knob.
6 Touch the Air temp. therapy control.
7 Use the rotary knob to set the air temperature value according to the
recommended air temperature range.
8 Press the rotary knob to confirm.
✓ The air temperature is set.
*
Sauer PJ, Dane HJ, Visser HK. New standards for neutral thermal environment of healthy very
low birth weight infants in week one of life.1984
Instructions for use Babyleo TN500 SW 1.0n
Recommended air temperature range