14.7.4.18 PAP (Pulmonary Artery Pressure Measurement)
To measure the items such as VPD (protodiastolic velocity) or VTD (telediastolic velocity)
in Spectral-Doppler mode:
14.7.4.19 Heart Rate (HR)
The method of the measurement is the same as that of measuring the Heart Rate in M mode.
operation review:
Heart Rate
®
Voluson
730Pro - Basic User Manual
105831 Rev. 0
1. Press the [Calc] key on the control panel.
The Cardiac calculations menu appears in the menu area.
2. Select [PAP] and [VPD] or [VTD] in the menu.
3. If necessary, select the [Angle] and the [Baseline].
4. Perform the measurement by using the trackball and the right or
left trackball key [Set].
(chapter 14.7.4.8)
Calculations and Patient Reports
14-41