5.3 Full 2-Port S-Parameter Measurements (F2P, optional)

The Full 2-Port mode (F2P) is an optional measurement mode for measuring
the frequency dependent S
is essential to perform a twelve-term calibration (12TC) before using this
mode (detailed information about performing a 12TC calibrations can be
found in chapter 4).
Fig. 9: Magnitude of S
Right-click an existing sample and select S
case the S
XY
further information.
A 12TC calibration must be activated for measuring S
indicated by three green LEDs in the Main Window.
Most of the functionality and handling is similar to the S
ments (display, result domains, masks, smoothing, data export, etc.),
please refer to section
1 S
is an abbreviation for S
XY
-parameters
XY
mode is not available, please
5.2
for further information on these topics.
, S
, S
, S
11
12
21
22
T3SP-Series Operator's Manual
1
for single-ended 2-port DUTs. It
- and S
-parameters
11
21
measurement mode. In
XY
Teledyne LeCroy support
for
-Parameters,
XY
-measure-
11
47