Exposure Calculation (Optional); Calculate - Teledyne CP Series User Manual

Table of Contents

Advertisement

User Manual
CP Series
Version 3.1 – June 2020
5.2
E
XPOSURE CALCULATION
Note:
This is an optional feature and will only be available when purchased.
5.2.1 C
ALCULATE
The exposure calculator software allows for automatic calculation of the
required kV or the required exposure time (T). The type of calculation
can be changed by pressing the 'OK' button (15) while the 'Calculate'
menu item is highlighted.
Enter the remaining desired variables and parameter required for the
calculation and press 'OK' on the 'Confirm' menu item to execute the
calculation.
The remaining items that must be completed to make the calculation
are:
[D]
When selecting the D(ensity) sub-menu you first must choose the
reference chart that will be used for the calculation. Charts are divided in
2 categories, ICM charts which are already pre-loaded in the PowerBox,
or Custom chart that customer can add manually.
(
)
OPTIONAL
 kV or TP: depending on the
 mA: the desired mA
 D: Select material, film type
 Thickness: the thickness of
 FFD: Film to Focal Distance
Exposure calc
Calculation
T
kV
200kV
mA
4.5mA
D Fe 2.0
AGFA D7
Thickness
35.0mm
FFD
700mm
**********

Calculate

selected calculation mode,
enter the desired kV or
Exposure time.
and desired quantity. See
below for further
explanation. Press 'OK' to
enter the sub-menu and
change the desired settings
the inspected material in
mm
in mm
Chart
ICM chart
Custom chart
38