11. Troubleshooting

Problem
After starting the software
the following warning
appears: "No valid
measurement modes
found!"
The red LED of the ON/OFF
button is blinking red.
The programs windows do
not show up or are
arranged strangely.
No calibration kit is
available for performing the
three-term calibration
(open, short, load)
T3SP-Series Operator's Manual
Approach
The measurement modes are read from
the device. Possibly the connection is not
established correctly. Make sure the
device is turned on (check if ON/OFF
buttons LED lights or blinks red).
If the message still appears, check the
Seunis.config file, if the entry Auto
connect is set to "true".
This is not an error but means that the
battery is currently recharged. The LED
blinks green while the device is turned off
and the battery is charged.
Press CTRL-ALT-W to restore the
factory-default window arrangement.
No CKD-file containing calibration kit
coefficients can be found by the software.
Make sure it is copied to the programs
Config subfolder. Alternatively it can be
installed via the keyboard shortcut CTRL-I
or the menu entry File
Main Window.
Install in the
65
931776 RevB