Teledyne T3SP Series Operator's Manual page 46

Time domain reflectometer
Hide thumbs Also See for T3SP Series:
Table of Contents

Advertisement

Right-click an existing sample and select S
TTC must be activated for measuring S
green LED in the Main Window.
Measurements can be performed either on channel 1 or 2 (only on a
channel where TTC is activated).
It can be switched between the different result domains magnitude-,
phase, Smith-Chart and VSWR view. For this purpose use the menu
entry View
S-Parameter or CTRL-SHIFT-F11/F12.
Multiple samples can be plotted by selecting rows in the Main Win-
dow while pressing down the CTRL-key.
User-defined tolerance masks can be used for pass/fail analysis. Use
the Impedance Measurement Properties Window for loading a seg-
mented tolerance mask (see also section 5.1.4). An example tolerance
mask can be found in the Examples subdirectory.
Please note: masks are only applicable to magnitude and VSWR traces.
Smoothing of the traces can be activated via the Advanced tab in the
Options Dialog (see section 2.5.4).
The results can be exported to various formats (e.g. Excel, CSV, Mat-
lab, and Touchstone) via the menu entry File
The frequency axis can be adjusted by changing to the Result Visual-
ization Window and pressing the shortcut ALT-F or via the menu item
View
S-Parameter
T3SP-Series Operator's Manual
11
-parameters, indicated by a
11
Change Frequency.
measurement mode.
Export.
931776 RevB
46