Right-click an existing sample and select S
•
TTC must be activated for measuring S
green LED in the Main Window.
•
Measurements can be performed either on channel 1 or 2 (only on a
channel where TTC is activated).
•
It can be switched between the different result domains magnitude-,
phase, Smith-Chart and VSWR view. For this purpose use the menu
entry View
S-Parameter or CTRL-SHIFT-F11/F12.
•
Multiple samples can be plotted by selecting rows in the Main Win-
dow while pressing down the CTRL-key.
•
User-defined tolerance masks can be used for pass/fail analysis. Use
the Impedance Measurement Properties Window for loading a seg-
mented tolerance mask (see also section 5.1.4). An example tolerance
mask can be found in the Examples subdirectory.
Please note: masks are only applicable to magnitude and VSWR traces.
•
Smoothing of the traces can be activated via the Advanced tab in the
Options Dialog (see section 2.5.4).
•
The results can be exported to various formats (e.g. Excel, CSV, Mat-
lab, and Touchstone) via the menu entry File
•
The frequency axis can be adjusted by changing to the Result Visual-
ization Window and pressing the shortcut ALT-F or via the menu item
View
S-Parameter
T3SP-Series Operator's Manual
11
-parameters, indicated by a
11
Change Frequency.
measurement mode.
Export.