The green check-mark in the upper-right corner (PASS indicator) of the
Main Window indicates whether the measurement passed the toler-
ance check. Otherwise the cross (FAIL indicator) indicates that the
tolerance check failed. Additional LEDs (green/red) are located in the
status bar of the Result Visualization Window (only if exactly one sam-
ple selected in list).
Please note: two different options can be chosen to classify the measure-
ment results. In the strict mode, the measurement is classified as NOT PASSED
if the tolerance mask is violated at any point. In the non-strict mode, the
measurement is classified as NOT PASSED if the average impedance result is
out of the limits. The mode can be set in the Seunis.config file (see section
2.2) by setting option
•
The tolerance mask can be turned on/off by using the
the tool bar. This also effects displayed impedance results (average &
min./max. impedance) and the small LED indicators in the status bar.
The indicators are grayed out if the tolerance mask is switched off.
•
The trace shown in the Result Visualization Window can be chosen
between impedance and step response via the menu entry View
TDR.
•
The x-axis can be changed from time [ns] to a space [cm] visualization
via the View
Window. Make sure a Reference Measurement is performed to move
the zero point ( = 0 cm) to a well defined value.
Please note: the propagation velocity according to the object under test
must be set in the Impedance Measurement Properties Window (cp. section
5.1.3).
•
An automatic zoom can be activated by checking the Auto Zoom check-
"Pass-fail strict" either to "1" or "0" .
TDR
x-axis menu entry in the Result Visualization
T3SP-Series Operator's Manual
button in
37