Teledyne T3SP Series Operator's Manual page 40

Time domain reflectometer
Hide thumbs Also See for T3SP Series:
Table of Contents

Advertisement

Remark: the software displays the x-axis of the impedance trace vs. the travel-
ing time t[ns] or versus a spatial resolution L[cm]. The optional PDF-export of the
results always shows a spatial [mm] resolution. The software needs some of the in-
formation (explicitly the velocity or alternatively line-type and geometry) to convert
the traveling-time into spatial information. Furthermore the reference plane can be
set to the TDR probe tip, by performing a reference measurement (F6 or Extras
Reference Measurement).
The General Information group holds information about the objects
under test and about the user. The information is used for creating
impedance reports stored in the SQC-project file.
1.
In the Customer field information can be stored for each sample.
2.
The labeling of the next three fields (by default Part Descr., Part
No., Batch No.) can be configured by the user in the Seunis.con-
fig (see 2.2) file by altering the line(s) "LabReplace SetWin
(...)" . Changes can be made for each installed language sepa-
rately.
3.
The User ID is a permanent setting and cannot be altered here.
Changes must be made in the Options Dialog (see section 2.5.4,
click menu Extras
4.
The fields Measurement Mode and Channel are purely informa-
tive and cannot be changed here.
In the Material/Line specification group some useful information about
the measurement object can be handled:
The Material field is purely informative.
If the Line Type is set to Other the propagation Velocity can be
T3SP-Series Operator's Manual
Options).
40
931776 RevB

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3sp15dT3sp10d

Table of Contents