CTRL-ALT-W can be used. Any user-defined setting can be saved as
default via View
now can be used to restore the default arrangement (also available
after a software restart).
4.
Click the toolbar button
Operation Mode.
5.
Prepare an empty sample list by clicking File
dow. In case a three term correction (TTC) is currently loaded, a ques-
tion dialog is raised. Click NO to clear the existing TTC, in this example
the TTC shall not be considered.
6.
Select the upper left cell and type in a Sample Name. You may add
more informative details of your samples in the Description column.
Remark: it is possible to prepare lists in other spreadsheet applications and
to copy the data into the sample list (copy & paste).
7.
Now select the requested measurement mode. The easiest way to do
this is to right-click the sample and click Select Measurement Mode.
Select the measurement mode TDR/Wave Impedance. Alternatively
the MODE button in the toolbar can be used.
8.
In the Oscilloscope Window the length of the TDR trace can be chosen
via the pull-down menu Number of points. Standard values from 512
to 32768 are available. The used Sampling time is preset to:
10 ps .
T3SP-Series Operator's Manual
Set default Arrangement. The shortcut ALT-W
to make sure you are in the Sample-List
New in the Main Win-
931776 RevB
=
25