checked.
13.
Make sure the first row is still highlighted. The highlighted sample will
actually be measured. Best practice is to click the row number in this
first column.
14.
Measure the sample either by clicking the measure button
right-lower corner of the Main Window or by using the F5-key.
Please note: if the TDR signal is temporarily instable, measuring is prohib-
ited. Please make sure the signal is stable before measuring. Instabilities
usually occur when there is a bad contact between probe and the object un-
der test. Please check the contacting by means of the real-time trace in the
Oscilloscope Window.
15.
Now the TDR traces (all average measurements) are additionally dis-
played (red) in the Oscilloscope Window. Please note that these are
not displayed when the impedance trace is displayed instead of the
TDR trace.
Remark: the measurement data can be deleted by using the shortcut STRG-
D or via the main menu Edit
the TDR data, all other information is preserved. Complete rows can be re-
moved by using the menu entry Edit
rows will be deleted. Attention: an UNDO function is currently not avail-
able.
16.
You can zoom into the TDR trace by using the mouse or the toolbar
buttons below the diagram. Press the left mouse button within the
display screen and select the zoom-area or use the mouse wheel. To
reset the view to full-view double right-click within the plot.