Teledyne T3SP Series Operator's Manual page 19

Time domain reflectometer
Hide thumbs Also See for T3SP Series:
Table of Contents

Advertisement

box can be checked. Usually this option should be unchecked.
The last valid TTC is loaded automatically in case the corresponding
box is checked.
In the ESD auto-protection field the number of seconds is set after
which the device switches on the ESD-protection automatically.
If the ESD-auto-off time is set to 0 the ESD-protection is only turned
off during the measurements. This is the safest operation mode and
recommended to be used in unsafe environments.
Advanced tab:
Stop macro execution on failure: see section
measurements.
Enable frequency domain smoothing: this option can be used for de-
noising frequency domain results (S-parameter measurements). By
default the video bandwidth is set to 20MHz . The value can be set in
the configuration file Seunis.config.
Equalize transmission coefficients in F2P: only enabled if full 2-port S-
parameter measurement mode is available. If checked the mean value
of S12 and S21 parameters are calculated, in order to suppress asym-
metries of the measurement system. This option is only reasonable
for linear passive devices.
Suppress information in PDF-report: the user can suppress the output
of pass/fail information, dielectric constant, full PDF saving path, step
rise time, and signal velocity.
Set TTC-location: for multi-user systems the saving location for the last
known TTC/12TC can be modified. This circumvents problems with
T3SP-Series Operator's Manual
2.6.2
about macro mode
19
931776 RevB