saved via the button Save segmented Tolerance Mask to an XML file,
•
cleared via the button Clear segmented Tolerance Mask (in this case
the standard rectangular tolerance mask becomes active again).
Another option to generate a user defined (segmented) tolerance mask is to
use the integrated mask generator. Click the menu item Extras
pedance Mask in the Main Window and the software will guide you through
the creation process. The generated XML file is not applied automatically
and must be loaded afterwards.
5.2 (D)S11 - Measurement Mode
These measurement modes (single-ended & differential) are for acquiring
the frequency dependent reflection coefficients (S
and time domain signals are connected via the Fourier-Transform thus TDR
signals can be transformed into S
ing in combination with a Fast-Fourier-Transform (FFT). It is essential to ac-
tivate the TTC before using this mode (detailed information about applying
TTC can be found in chapter 4).
Fig. 8: Magnitude and Smith-Chart of reflection coefficient S
T3SP-Series Operator's Manual
11
by applying a proper signal preprocess-
11
Create Im-
-parameter). Frequency-
11
931776 RevB
45