Sirona Dental Systems GmbH
Care and cleaning instructions for the practice team C2
3
Vacuum system
3.1
Care
3.2
Disinfection with a dry suction system
8
A
59 58 512 D 3370
D 3370.103.01.13.02
+
+
, M1
After each patient
To ensure that the vacuum system is always ready for
use, you must aspirate a large glass of cold, clear water
through the suction hoses used after each patient,
especially after each intervention during which the
patient loses blood and after using hydrogen peroxide
(H
O
).
2
2
During prolonged treatments, you must aspirate a
glass of water at least every 60 minutes.
Daily during the noon break and at the end of
each workday
The vacuum system is subjected daily to septic secre-
tions, saliva and blood. For hygienic reasons the system
must therefore be disinfected with a detergent/disinfec-
tant approved by Sirona.
ATTENTION
Use only cleaning and disinfecting agents which have
been approved by Sirona (see page 2, General informa-
tion)!
Do not use any washing powder or other disinfec-
tants!
•
Prepare the cleaning solution in a separate contain-
er according to the manufacturer's instructions and
mix thoroughly.
•
Remove the gold trap.
•
Clean the cuspidor.
•
Plug in the cleaning adapter up to the stop. For the
saliva ejector hose, insert the intermediate piece (8)
into the cleaning adapter.
•
Fill the cleaning solution up to the filling limit (A) in
the cuspidor (approx. 0.75 liters)
•
Remove the suction hoses from their holders and at-
tach them to the side of the cleaning adapter, if pos-
sible all at once.
•
The cleaning solution is aspirated.
Observe manufacturer's instructions regarding the
reaction time!
3 Vacuum system
3.1 Care
.
7