Performance Test Record (Page 1 Of 3) - Agilent Technologies 53147A Assembly Level Service Manual

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

Chapter 1 Performance Tests

Performance Test Record (Page 1 of 3)

Performance Test Record (Page 1 of 3)
Agilent 53147A/148A/149A Frequency Counter/Power Meter/DVM
Serial Number:
Test Performed By:
Date:
Notes:
Pre Calibration Test
Test
Number
Operational Verification
1
Power on Self Tests
2
10 MHz Test
3
Run Self Test
4
10 MHz External Timebase Input Test
5
Power Meter Test
6
DVM Test
Test
Number
Performance Tests
Channel 1 Frequency Sensitivity
1a
10 Hz
30 Hz
1 kHz
500 kHz
5 MHz
10 MHz
20 MHz
Channel 1 Frequency Sensitivity
1b
50 MHz
125 MHz
Assembly Level Service Guide
Complete
Repair/Work Order No.
Temperature:
Relative Humidity:
Post Calibration Test
Specifications
N/A
10,000,000 Hz
N/A
2 MHz
.
0 dBm ±
02 dB
+25 Vdc ±0.25% ±10 mV
(+24.365 Vdc to +25.635 Vdc)
Specifications
10 Hz @ 40 mV rms (–14.9 dBm)
30 Hz @ 40 mV rms (–14.9 dBm)
1 kHz @ 25 mV rms (–19.2 dBm)
500 kHz @ 25 mV rms (–19.2 dBm)
5 MHz @ 25 mV rms (–19.2 dBm)
10 MHz @ 25 mV rms (–19.2 dBm)
20 MHz @ 25 mV rms (–19.2 dBm)
50 MHz @ 25 mV rms (–19.2 dBm)
125 MHz @ 25 mV rms (–19.2 dBm)
Test Results
Pass
Fail
Test Results
Pass
Fail
1-39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53148a53149a

Table of Contents