Chapter 1 Performance Tests
Operational Verification
To Reference
10 MHz Connector
(Rear Panel)
Agilent
53147A/48A/49A
Agilent
Figure 1-1. External Timebase Test Setup
4
Cycle the POWER button to preset the instrument.
5
Press the Shift key on the instrument's front panel.
6
Press the Menu key on the instrument's front panel.
7
If necessary, use the up- and down-arrow keys to cycle the display until
it reads REF OSC> INT.
8
Press the right-arrow key once. INT begins to flash off and on.
9
Press the up- or down-arrow key once. INT changes to EXT.
10
Press the Enter key.
11
Press the Chan Select key once to select the CHANNEL 1 input connector.
Assembly Level Service Guide
Agilent 3325A/B Synthesizer
Channel 1 (with 10100C 50Ω Feedthrough Attached)
Agilent 3325A/B Synthesizer
10 MHZ Output
2 MHz Output
1-11