Chapter 2 Service
Calibration Procedures
14
When the calibration program displays:
Update from old 10974.01 to new 10974.96 Cal. Constant? (y/n) y
type "y" (the values in the display line above are simulated).
15
When the calibration program redisplays its menu (as shown in
Figure 2-8), type "e" to exit the program.
16
Close the terminal program and disconnect the serial cable from the
computer and the instrument.
2-22
Assembly Level Service Guide