Chapter 1 Performance Tests

Operational Verification

Verify that the front-panel display shows the following:
All segments of the front-panel display are temporarily activated.
TESTING is displayed.
SELF TEST OK is displayed.
The model number of the instrument is displayed (a four-digit hex
number is also displayed).
GPIB ADDR nn (nn = a two digit number from 0 to 30) is displayed.
CH2 NO SIGNAL is displayed.
If an error message is displayed, refer to the troubleshooting section in
Chapter 2, "Service."
Mark Pass or Fail in the Performance Test Record on page 1-39, Test 1.
To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment and the
timebase to stabilize.