Agilent Technologies 53147A Assembly Level Service Manual page 122

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

Chapter 6 Specifications
Introduction
Input Characteristics (continued)
Emissions (typical)
("kickback noise")
Channel 1
Channel 2 (measuring/no input)
Resolution selection
Channel 1/Channel 2
Accuracy
Channel 1/Channel 2
(LSD=Resolution selected)
Residual Stability*
Channel 1
Channel 2
*Counter and source tied to same timebase
Measurement Time (typical)
Channel 1
Channel 2
Acquisition Time (typical)
(1 MHz FM rate, power meter off)
Channel 1
Channel 2 (FM Auto/FM Off)
FM Tolerance
Channel 1
Channel 2 (FM Auto)
(FM Off)
AM Tolerance
Channel 1, Channel 2
Amplitude Discrimination
Channel 1
Channel 2
below 300 MHz
above 300 MHz
Timebase
Frequency
Output
External timebase input
Connector
Internal timebase stability
Aging rate per day
Aging rate per month
Short term (1 sec. average time)
Line variation (± 10%)
Warm-up
Temperature stability (0-55ºC)
6-4
53147A
N/A
−40 dBm/<−70 dBm
1 Hz to 1 MHz
±1 LSD
± residual stability
± timebase error x frequency
N/A
0.6 LSD rms
1/Resolution +30 ms
1/Resolution +
Acquisition time + 30 ms
N/A
150 ms/125 ms
N/A
20 MHz p-p max
@ 10 MHz rate
1 MHz p-p @ 10 MHz rate
Any index provided minimum
signal level is not less
than sensitivity
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
10 MHz
10 MHz sine wave, 1 Vrms into 50Ω
1, 2, 5, 10 MHz, 1 to 5 Vrms into 50Ω
BNC female located on rear panel
7&;
(standard)
N/A
-7
<1 X 10
-9
<1 X 10
-7
<1 X 10
N/A
-6
<1 X 10
53148A
N/A
−40 dBm/<−70 dBm
1 Hz to 1 MHz
±1 LSD
± residual stability
± timebase error x frequency
N/A
0.8 LSD rms
1/Resolution +30 ms
1/Resolution +
Acquisition time + 30 ms
N/A
150 ms/125 ms
N/A
20 MHz p-p max
@ 10 MHz rate
1 MHz p-p @ 10 MHz rate
Any index provided minimum
signal level is not less
than sensitivity
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
2YHQ
(Option 001)
-10
<5 X 10
-8
<1.5 X 10
-10
<2 X 10
-10
<1 X 10
-8
<1 X 10
within 5 min. after turn-on at 25ºC
-8
<1 X 10
Assembly Level Service Guide
53149A
N/A
−40 dBm/<−70 dBm
1 Hz to 1 MHz
±1 LSD
± residual stability
± timebase error x frequency
N/A
1.25 LSD rms
1/Resolution + 30 ms
1/Resolution +
Acquisition time + 30 ms
N/A
165 ms/140 ms
N/A
20 MHz p-p max to 26.5 GHz,
12 MHz p-p max above
26.5 GHz @ 10 MHz rate
1 MHz p-p @ 10 MHz rate
Any index provided minimum
signal level is not less
than sensitivity
N/A
N/A
Automatically measures the largest
signal present provided signal is >10
dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for: