Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
2
Connect the free end of the Agilent 11730A Power Sensor Cable to the
input connector on the Agilent E4418B Power Meter.
3
Connect the Agilent 8481A Power Sensor to the power reference output
connector of the Agilent E4418B Power Meter.
4
Zero the E4418B Power Meter.
5
Calibrate the E4418B Power Meter.
6
Disconnect the Agilent 8481A Power Sensor from the Agilent E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector
on the Agilent 53147A/48A/49A (see Figure 1-9).
Agilent
53147A/48A/49A
Agilent
Figure 1-9. Power Meter Reference Level Test Setup
7
Record the power reading shown on the Agilent E4418B Power Meter for
Test 5 in the Performance Test Record on page 1-41.
8
Mark Pass or Fail for Test 5 in the Performance Test Record.
Assembly Level Service Guide
Output connector (with Agilent 8481A
Power Sensor Attached)
Agilent E4418B
Power Meter
Input
1-33