Chapter 2 Service
Calibration Procedures
Calibration Access Hole
Figure 2-1. Calibrating the 10 MHz Frequency Reference Oscillator
4
Insert a non-metallic, slot-type adjustment tool through the hole near the
bottom of the instrument's left side and into the slot in the timebase
calibration screw.
5
Turn the calibration screw clockwise to increase the Counter's reading or
counter-clockwise to decrease the reading until the frequency value
displayed on the Counter's front panel is exactly 10,000,000 Hz.
2-8
Assembly Level Service Guide