Performance Test Record (Page 3 Of 3) - Agilent Technologies 53147A Assembly Level Service Manual

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

Chapter 1 Performance Tests

Performance Test Record (Page 3 of 3)

Performance Test Record (Page 3 of 3)
Test
Complete
Number
Performance Tests
Power Measurement:
3
3 µW
10 µW
30 µW
100 µW
300 µW
1 mW
3 mW
10 mW
30 mW
100 mW
20.00 dBm
–10.00 dBm
Power Reference Frequency:
4
50.0 MHz
Power Reference Level:
5
1.0 mW (0.00 dBm)
Assembly Level Service Guide
Specifications
± 0.02 dB or ± 0.5%
"
"
"
"
"
"
"
"
"
±0.04 dBm
±0.04 dBm
50.0 MHz ± 1%
(49.5 MHz to 50.5 MHz)
1.0 mW ± 1.2%
0.988 mW to 1.012 mW
Test Results
Pass
Fail
1-41

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53148a53149a

Table of Contents