Chapter 3 Replacing Assemblies

Removing the Handle

Removing the Handle
Use the following procedure to remove the carrying handle:
1
Remove the front bumper (see "Removing the Bumpers" on page 3-4).
2
Stand the instrument on end on its back bumper.
3
Position the handle so that it is against the top surface of the shroud
(see Figure 3-2).
Figure 3-2. Removing the Handle
4
Use a T15 TORX screwdriver to remove two of the 8-32 x 1/4-inch handle
mounting screws (remove the two screws that are closest to the back of
the instrument).
5
Repeat Step 4 for the other handle mount.
6
While holding the handle to support it, remove the remaining two screws
(one on each handle mount).
7
Remove the handle.
Assembly Level Service Guide
3-5