Agilent Technologies 53147A Assembly Level Service Manual page 50

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

NOTE
22
23
24
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
The following is an example showing how a calculation of the Power
Reference Output Level is performed:
R = 200
V 0 = 170 µV or 170 x 10
V 1 = 78,107 µV or 78,107 x 10
V comp = 5.1583 V
R = previously recorded value
CalibrationFactor = 99.68%
2 V
Prf =
Note that V
is eventually dropped out of the equation in the following
0
steps, because its value is very small (effectively zero).
2 (5.1583) (.078 – 0)
P rf =
0.80469 – 0.00608
=
= 1.00147 mW
Verify that P rf is between 0.988 mW and 1.012 mW.
Record the value for P rf in the Performance Test Record on page 1-41.
Mark Pass or Fail for Test 5 in the Performance Test Record.
1-38
–6
V
–6
2
( V
– V
) + V
– V
comp
1
0
0
4R (Calibration Factor)
2
+ (.000170)
4(200) (99.68%)
0.79861
=
797.44
797.44
Assembly Level Service Guide
or .078 V
2
1
2
– (0.78)