Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
Repeat steps 5, 6, and 7 for each of the remaining frequencies specified for
Test 1a. Don't forget to change the power level when performing the
1 KHz test (–19.2 dBm/25 mV).
If all of the frequency values you recorded for Test 1a are correct, mark
Pass in the Performance Test Record for Test 1a. If any of the frequency
readings you recorded are incorrect, mark Fail in the Performance
Test Record.
Do not disconnect the cable from the instrument's Reference 10MHz
connector. This connection is used in the following test.
Do not turn off the instrument. Turning the instrument off and back on
resets the reference oscillator (REF OSC) selection to internal (INT)
and the channel selection to CHANNEL 2.

Test 1b: 50 MHz to 125 MHz Frequency Sensitivity

Connect an Agilent 10100C 50 Ω Feedthrough Terminator to the
CHANNEL 1 input connector on the instrument's front panel (refer to
Figure 1-4 for steps 1 through 9).
Agilent 83650B Synthesizer
Output
To Reference 10 MHz
Agilent
53147A/48A/49A
Agilent
Figure 1-4. 50 MHz to 125 MHz Frequency Sensitivity Test Setup
1-20
RF Output
Connector
(Rear Panel)
Agilent
10100C 50
Feedthrough
Agilent
Agilent
11904A 2.92 mm to 2.4 mm Adapter
Agilent
1250-2015 BNC to SMA Adapter
Channel 1
Assembly Level Service Guide
Agilent 437B Power Meter
Input
Agilent
11667C
Power Splitter
Agilent
8487A/D
Power Sensor
8490D 20 dB Attenuator