Agilent Technologies 53147A Assembly Level Service Manual page 26

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

Chapter 1 Performance Tests
Operational Verification
5
Press the Zero key.
The display shows ZEROING for several seconds.
6
Connect the power sensor head to the Power Meter OUTPUT connector.
7
Press the Cal key.
The display briefly shows CALIBRATING.
8
Turn on the power reference output in the instrument's menu
(PWR REF > ON).
9
Verify that the Counter displays a reading of 0 dBm ±.02 dB.
10
Mark Pass or Fail for Test 5 in the Performance Test Record on page 1-39.
11
Disconnect the test setup.
1-14
Assembly Level Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53148a53149a

Table of Contents