Chapter 1 Performance Tests
Operational Verification
12
Set the frequency of the first Synthesizer (connected to the instrument's
rear-panel Reference 10MHz connector) to 10 MHz at 1 V rms.
13
Set the frequency of the second Synthesizer (connected to the instrument's
Channel 1 input connector) to 2 MHz at 100 mV rms.
14
Press and release the instrument's Reset/Local key.
15
Verify that the Counter displays a reading of approximately 2 MHz and
that the Ext Ref annunciator is activated.
16
Mark Pass or Fail for Test 4 in the Performance Test Record on page 1-39.
17
Disconnect the test setup.
1-12