Chapter 1 Performance Tests
Equipment Required
Equipment Required
Table 1-1 lists the test equipment and accessories needed to perform the
tests in this chapter.
Table 1-1. Recommended Test Equipment and Accessories
Instrument Type
Synthesized Signal
Generator
Microwave Synthesized
Signal Generator
Microwave Amplifier
Digital Multimeter
DC Power Supply
Frequency Counter
Power Meter
Power Meter
Power Meter
Thermistor Mount
Power Sensor
Power Sensor
Power Sensor Cable
Power Sensor Cable
Range Calibrator
* Equivalent equipment can be substituted for all instruments and accessories.
** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
† One supplied with Agilent 53147A/48A/49A
1-4
Required Characteristics
10 Hz to 20 MHz, +7 dBm to –40 dBm
10 MHz to 50 GHz, Accuracy >1 x 10
+7 dBm to –50 dBm
45 MHz to 50 GHz
6½ digit AC/DC
25 VDC
10 Hz to 125 MHz
Range 1 mW
50 MHz to 50 GHz, +13 dBm to –50 dBm
100 kHz to 110 GHz, –70 dBm to +44 dBm
SWR 1.05 at 50 MHz
Accuracy ±0.5% at 50 MHz
10 MHz to 18 GHz, -30 dBm to +20 dBm
50 MHz to 50 GHz, +13 dBm to –50 dBm
2.4 mm to 2.4 mm, 5 ft.
3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
Model
Recommended *
Agilent 3325A/B
-10
Agilent 83650B
Agilent 83051A
Agilent 34401A
Agilent E3640A
Agilent
53150A/51A/52A
Agilent 432A
Agilent 437B
Agilent E4418B
Agilent 478A-H75
or 478A-H76
Agilent 8481A
Agilent 8487A
Agilent 11730A
Agilent 8120-1082
Agilent 11683A
Assembly Level Service Guide
Quantity
Required Use **
2
OV, P, T
1
P, T
1
P, T
1
OV, P
1
OV
1
P, C, T
1
P, C, T
1
P, C, T
1
P, C, T
1
P, T
1
P, T
1
P, C, T
†
2
P, T
1
P, T
1
P