Chapter 3 Replacing Assemblies

Removing the Power Supply

Removing the Power Supply
Use the following procedure to remove the power-supply:
1
Remove the shroud (see "Removing the Shroud" on page 3-7).
2
Place the rear bumper face-down on your work surface (the portion of the
bumper that normally fits around the end of the instrument should be
facing up).
3
Carefully position the instrument on end with the front panel inserted
into the bumper.
4
Disconnect the two power-supply connectors (one at the front and one
at the rear of the power supply).
5
While holding the power supply in position, use a #2 Phillips screwdriver
to remove the four countersunk flathead screws that fasten the power
supply to the side wall of the chassis (see Figure 3-8).
Figure 3-8. Removing the Power Supply
3-14
Assembly Level Service Guide