Chapter 1 Performance Tests
Operational Verification
Run Self Test
This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
Disconnect all signal cables from the input and Reference 10MHz connectors.
Press and release the Shift key, and then press and release the Menu
(Reset/Local) key.
Press the up- and/or down-arrow key(s) as many times as necessary until
DO SELF TEST is displayed.
Press and release the Enter key.
If no errors are detected, SELF TEST OK is briefly displayed when the self
test is completed. If any error messages are displayed, refer to the
troubleshooting section in Chapter 2, "Service."
Mark Pass or Fail in the Performance Test Record on page 1-39, Test 3.
Assembly Level Service Guide