Eac Explosionproof - Oval ALTI mass CA00A Instructions Manual

Table of Contents

Advertisement

L -- 740 -- 28 -- E

10.4.5 EAC explosionproof

o TC RU C-JP.Γb05.B.00629
N
Applicable standard:
● Integral type
Converter symbol:
Detector symbol:
Ambient temp.:
Detector to be connected: CA006 to CA150
Fluid temp.:
Power source:
Allowable voltage:
Communication:
● Separate type
Converter symbol:
Detector symbol:
Converter Ambient Temp.: -40℃ to +55℃
Detector Ambient Temp.: As shown on Detector Model Specification Table 2
Fluid Temp.:
Power source:
Allowable voltage:
Communication:
Explanation about subject model codes
Model code: C①②③④-x⑤xxx-5⑥x-1⑦⑧x⑨x-x
Position
Meaning
Sensor Series
Nominal Size
②③④
Temperature
Classification
Temperature Class
Converter Installation
Structure
Power Source
Communication
Configuration
1
ΓOCT 30852.0-2002 (MeK 60079-0:1983)
ΓOCT 30852.1-2002 (Mek 60079-1:1998)
ΓOCT 30852.10-2002 (Mek 60079-11:1999)
1 Ex d ib ⅡC T4X (When connected with CA series)
1 Ex d ib ⅡB T4X (When connected with CB and CS series)
1 Ex ib ⅡC T4 (CA series)
1 Ex ib ⅡB T4 (CB, CS series)
-40℃ to +55℃ (-20℃ to +55℃ when connected with CS080 and CB040/050)
CB006 to CB050
CS010 to CS080
As shown on Detector Model Specification Table 2
AC100 to 240V 50/60Hz, or DC20 to 30V
Um=250V
HART, Modbus, PROFIBUS or FOUNDATION Fieldbus(FISCO)
1 Ex d[ib] ⅡC T6X
1 Ex ib ⅡC T1, T2, T3, T4, T5, Gb (CA series)
1 Ex ib ⅡB T3, T4 (CB, CS series)
As shown on Detector Model Specification Table 2
AC100 to 240V 50/60Hz, or DC20 to 30V
Um=250V
HART, Modbus, PROFIBUS or FOUNDATION Fieldbus(FISCO)
Value
A
CA series
B
CB series
S
CS series
CA00A, CA001, CA003, CA006, CA010, CA015, CA025, CA040, CA050,
00A to 250
CA080, CA100, CA150, CA15H, CA200, CA20H and CA250
006 to 050
CB006, CB010, CB015, CB025, CB040 and CB050
010 to 080
CS010, CS015, CS025, CS040, CS050, CSR50 and CS080
1
Standard(-40℃ to +130℃)
2
Standard(+130℃ to +200℃)
3
High temp.(+200℃ to +350℃)
4
Low temp.(-200℃ to +50℃)
1
T1: Applicable sensor sizes CA025 to CA150
2
T2: Applicable sensor sizes CA003 to CA250
3 or 4
T3 or T4: All sensor sizes applicable
5
T5: Applicable sensor sizes CA025 to CA250
1
Integrally mounted models
2
Separately mounted models
1
20 to 30VDC
2
85 to 264VAC (Safety Rating: 100 to 240VAC) 50/60Hz
1
HART Communication (HART Protocol Version 7, Bell 202)
2
FOUNDATION Fieldbus H1 Communication (ITK Version 6)
3
PROFIBUS PA Communication (Profile Version 3.02)
4
Modbus Communication (RS-485 Modbus Protocol)
Description

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents