10.3 Replacement Parts

We recommend to keep a stock of replacement parts in case anything goes wrong. Replacement parts
are available at the OVAL representative in your area or OVAL Customer Service Department. Please
supply us with the flowmeter model, part name, product number, and quantity when you order.
Interconnect cable (Teflon or PVC): Quantity 1
Replacement of parts such as o-rings and sheet packing for transmitter and
terminal box
O-rings and gaskets are used as sealing components with limited service life.
Although the exact service life depends on the operating environment, periodic replacement is
recommended. Recommended replacement period: 5 to 8 years after the delivery
CAUTION:
1. Whenever you open or disassemble the instrument, inspect the O-rings
and gaskets. Replace the deteriorated parts.
2. To ensure the soundness, periodic re-torquing of the corresponding parts
is recommended.
Corresponding parts:
Transmitter: O-ring between display lid (front) - body
Transmitter: O-ring between terminal box lid (rear) - body
Transmitter: Sheet packing between side lid - body
Transmitter: O-ring between adopter - body (Integral type only)
Transmitter: O-ring on sensor unit (Integral type only)
O-ring between terminal box - lid (Separate type only)
Cable gland: rubber seal and O-ring
Please contact the OVAL sales office when replacement is necessary