6.10 How to Change Transmitter Orientation

WARNING: Be sure to turn off power before you start working.
To change transmitter orientation, follow the steps given below:
(1) Turn off power.
(2) Make preparations for conduit and cable so that they will by no means cause you any trouble while
changing transmitter orientation.
Work on a level plane with the transmitter up and the sensor secured so that loosening bolts will not
cause the transmitter to come off.
(3) Loosen bolts holding the transmitter adapter with hex wrench. See Fig. 4.10 (M8 hex socket head
screws at four places).
WARNING: Never loosen the hex socket head screws located on the
WARNING: You may loosen bolts but never separate the transmitter from
(4) Rotate the transmitter to the desired position and secure it with setscrews (four places).
Loosen four hex socket head screws (M8) at the
adapter and change the transmitter orientation.
Never loosen these screws
located on the transmitter body.
CAUTION
The transmitter may be rotated over an 180° arc as shown in Fig. 6.28. But rotating it
beyond 180° will twist the harness from the sensor unit to the extent the equipment is
damaged.
(5) Confirming that the transmitter is secured in place, make conduit and wiring connections.
(6) Verify that the flowmeter operates properly.
transmitter body.
the sensor unit.
Transmitter rotatable over an arc of 180° max.
M8 Setscrew
(4 places)
Fig.6.28