10.5 Ambient Temperature (Type U)

Allowable ambient temperature permitted for the sensor unit is as described in the table below.
(The following table describes the condition for the non-explosionproof models. For the explosionproof
models, make sure to satisfy the temperature conditions described in Section 10.4, as well as the
condition described below.)
Sensor unit temp. class
Standard (200℃ and lower)
High temperature service model
(350℃ and lower)
Low-temperature explosionproof
service model
Separate type sensor unit (temp. class: standard) Max. ambient temp. range
65
60
55
50
45
40
35
30
25
20
15
10
5
0
−200
Please contact us in the case the ambient temperature exceeds max. ambient temperature in the graph.
NOTE:
Terminal box
Adapter
1
[Fluid temperature] 150℃ and lower
[Ambient temperature] −40 to +55℃
−150
−100
−50
Heat insulation can be applied to
the sensor unit itself at the fluid
temperature. However, do not apply
heat insulation to terminal box or
adapter.
Transmitter construction
Integral type
None
None
0
50
Fluid temperature (℃)
MAX. ambient temperature
Separate type
[Sensor unit ambient temp.]
−40℃ to max. ambient temp.
in the graph below
[Sensor unit ambient temp.]
−40 to +55℃
[Sensor unit ambient temp.]
−20 to +50℃
100
150
200