Status Output - Oval ALTI mass CA00A Instructions Manual

Table of Contents

Advertisement

L -- 740 -- 28 -- E

9.7.3 Status output

A status output is available at any level desired.
This section covers the procedure to provide a simulated status output in the setting ON. A similar
procedure applies to other output settings
Click "SEL1".
Click "SEL3".
Click "SEL3".
Click "SEL3".
Click "ENT".
Click "SEL3".

(1) Bring up the item of function you want to use.
Firstly, proceed key operation to show a screen for simulated output of status
output (Status Output).
(For the procedure to view individual items, see 9.2.2 thru 9.2.7 Transition
chart.)
When the cursor is placed over Status Output, click "SEL1" to bring up
simulated output setup screen for status output.
(2) Decide whether to activate the function or not.
If you want to start simulated output of status output, click "SEL3" to proceed
to the next screen.
To abort the procedure without running a simulated output, click "ENT" and
then "SEL3". The window returns to menu select screen.
Click "ENT".
Click "ENT".
(3) Enter a simulated output value.
Select the simulated output mode.
For the setup procedure, see 9.4 Parameter Selection (3).
(4) Run a simulated output.
If setting "On" is definitely OK, click "SEL3".
If you want to abort the simulated output process, click "ENT" and then click
"SEL3".
The window returns to menu select screen.
Click "ENT".
Click "ENT".
(5) Status output is now generated.
Output of status output is fixed to On.
(Unaffected by obtained measurements, error processing or other factors.)
(During the check, GRN LED blinks on and off.)
If you want to cancel the simulate output process, click "SEL3".
Simulated output stops and the window returns to initial menu select screen.
Click "ENT".
Click "SEL3".
Click "SEL3".