(3) "SEL3" operation
This operation is used to finalize the numeric value that has been changed and activate it, etc.
Here we cover "SEL3" operation as an example of switching to MODE SELECT menu screen.
9.1.3 View the variables
(1) Description of displayed information
Shown below is a flow diagram with CA sensor (default settings).
At right of each screen is the description of information displayed in the screen.
The window changes from one screen to another with switch operation. Repeat touching the switch
until the desired screen appears.
Click "ENT".
Click "ENT".
(※1 with ↓)
(※1 with ↑)
Click "ENT".
Click "ENT".
(※1 with ↓)
(※1 with ↑)
Click "ENT".
Click "ENT".
(※1 with ↓)
(※1 with ↑)
① Touch SEL key at lower left of screen over glass faceplate with
② Hold you finger on.
③ While holding your finger on, the screen repeats a change between
④ End of "SEL3" operation
Mass Flow: Instant mass flowrate
......
Decimal point placement is alterable by config. (See 9.2.2 Transition chart (1) [No. 1-3-4-1].)
Units of measure are alterable by config. (See 9.2.2 Transition chart (2) [No. 2-1-1].)
Vol Flow or Vol Flow(s): Instant volume flowrate
......
Decimal point placement is alterable by config. (See 9.2.2 Transition chart (1) [No. 1-3-4-2].)
Units of measure are alterable by config. (See9.2.3 Transition chart (2) [No. 2-1-2].)
If fixed density is your option, Vol Flow(s) appears.
Below 0.3g/m in density, volume flow is clamped to 0.
Density: Fluid density
......
Decimal point placement is alterable by configuration. (See 9.2.2 Transition chart (1) [No. 1-3-4-3].)
Units of measure are alterable by config. (See 9.2.3 Transition chart (2) [No. 2-2-1].)
Temp: Fluid temperature
......
Units of measure are alterable. (See 9.2.2 Transition chart (2) [No. 2-3-1].)
Counter 1: The number of counts of pulse output 1.
......
When reading exceeds "99999999", it resets to 0.
The reading is retained irrespective of power cycling.
Can be reset to 0 with Count/Total 1 reset function. (See 9.2.7 Transition chart (6) [No.3-5-1].)
If pulse output 1 related parameter is modified, the counter is reset to 0.
......
Counter 2: The number of counts of pulse output 2.
When reading exceeds "99999999", it resets to 0.
The reading is retained irrespective of power cycling.
Can be reset to 0 with Count/Total 2 reset function. (See 9.2.7 Transition chart (6) [No. 3-5-2].)
If pulse output 2 related parameter is modified, the counter is reset to 0.
your finger.
When the key responds, a letter S appears in the highlighted S3
area at lower right of screen.
(On certain screens, letter S3 may appear at lower center of
screen.)
After letter S comes on for 0.5 sec, the letter S appears highlighted
as shown in the following screen.
highlighted and unhighlighted.
On the third appearance of highlighted S (duration 0.5 sec),
remove your finger. A switch to MODE SELECT menu screen
takes place.
A switch to MODE SELECT menu screen is complete.
L -- 740 -- 28 -- E