How To Change Transmitter Display Orientation - Oval ALTI mass CA00A Instructions Manual

Table of Contents

Advertisement

L -- 740 -- 28 -- E
WARNING: With CS sensor unit, do not attempt to adjust orientation with hex
socket head bolts shown in the figure below, or its functions as a
flowmeter will be lost.
Do not remove socket
head bolts.

6.11 How to Change Transmitter Display Orientation

The transmitter display may be rotated through 360 in steps of 90° within the housing.
WARNING: Be sure to turn off power and discharge your static electricity before you
work.
① Remove the locking part.
Fig.6.30

Fig.6.29
② Using a flat tool 10mm wide
approx., turn the display lid
counterclockwise
to
and loosen further by hand. (Be
careful so as not to damage the
finish.)
Fig.6.31
③ Disengage the hooks (3 places).
(Recommended tool: Precision
loosen,
flathead screw-driver)
Fig.6.32