9.12.4 Double pulse output
This function provides pulse output 2 in sync with pulse output 1.
It is enabled by setting up one of the following to pulse output 2 assignment.
(1) Setup procedure
① Pulse output 2 assignment (Assign)
・Double pulse 90: Double pulse output 90°
Pulse output 2 lags pulse output 1 by 90 degrees (1/4 of the period) when it goes into action.
(The maximum lag is 1 msec, above which is a lag within 1/4 of the period, however.)
・Double pulse -90: Double pulse output -90°
Pulse output 1 lags pulse output 2 by 90 degrees (1/4 of the period) when it goes into action.
(The maximum lag is 1 msec, above which is a lag within 1/4 of the period, however.)
・Double pulse 180: Double pulse output 180°
Pulse output 2 goes into action, polarity of pulse output 1 waveform inverted.
(2) Examples of setting and behavior
① Double pulse 90°
・Fast pulse output (above 250Hz)
Pulse out 1
Pulse out 2
・Slow pulse output (below 250Hz)
Pulse out 1
Pulse out 2
② Double pulse -90°
・Fast pulse output (above 250Hz)
Pulse out 1
Pulse out 2
・Slow pulse output (below 250Hz)
Pulse out 1
Pulse out 2
③ Double pulse 180°
Pulse out 1
Pulse out 2
 90
(1/4 of the period)
1msec 
90
(1/4 of the period)
1msec        