Pulse Output Wiring; Status Output Wiring - Oval ALTI mass CA00A Instructions Manual

Table of Contents

Advertisement

7.4 Pulse Output Wiring

When pulse output is to be connected to a receiving instrument, Pulse Output 1 has terminals P1 (
terminal P1 (
) while Pulse Output 2 has terminal P2 (
If voltage pulse signal is your option, signal transmission length has restrictions on output frequency.
※ Values are based on wire material 0.75 sq in cross section.
To change from voltage pulse output to open drain pulse output, or vice versa, remove the front lid and
select jumper positions as indicated below.
Pulse output setup procedure appears in 9.7.2 Analog output.
※ Pulse output will be invalid if FOUNDATION Fieldbus, PROFIBUS PA is selected as communication
interface.

7.5 Status Output Wiring

Status output appears across terminals S.O. (
Status output setup procedure appears in 9.7.3 Status output.
※ Status output will be invalid if FOUNDATION Fieldbus, PROFIBUS PA is selected as communication
interface.
Output frequency
10kHz
1kHz
100Hz
) and S.O. (
) and terminal P2 (
Max. transmission length ※
10m
100m
1000m
JP3
JP4
Fig.7.7
).
L -- 740 -- 28 -- E
).
Pulse output 1 jumper
OPEN: Open drain pulse
CLOSE: Voltage pulse
Pulse output 2 jumper
OPEN: Open drain pulse
CLOSE: Voltage pulse