Agilent Technologies 6850 Series Service Information page 73

Gc system

Advertisement

5. Disconnect the sampling valve from the inlet
at valve port 1. Cap the line to the inlet using
the tube cap from the direct-tube assembly.
Install the direct-to-column line
1. Open the lid and disconnect the column from
the inlet. Plug the inlet column fitting with a
no-hole ferrule and column nut.
Caution
Do not use the 1/16-inch direct-tube assem-
bly supplied in this kit to punch the hole
through the insulation unless the protective
cap is on. If left unprotected, insulation will
be trapped in the tubing!
Released: April 2004
2. Locate the small through hole in the oven lid
Figure 1-18. Location of oven lid through
6850 Series Gas Chromatograph Service Procedures
between the purged packed inlet and the
valve box. There is a corresponding hole in
the bottom of the oven lid, behind the inlet.
Using a spare 1/16-inch tube, a wire, or a
thin screwdriver, clear the insulation from
this hole. See
Figure
1-18.
(Liquid sampling valve shown)
hole
Oven lid
through
hole
73