Agilent Technologies 6850 Series Service Information page 225

Gc system

Advertisement

7. Remove the fan from the instrument.
8. Disconnect the fan wires at the connector.
When reinstalling the fan, orient the electrical
wires to the front and down. The airflow indica-
tor arrow should point up.
Released: April 2004
Injection port cooling fan
1. Turn off the instrument and remove the
2. Remove the top cover. See
6850 Series Gas Chromatograph Service Procedures
WARNING
Shock hazard. Before beginning, turn off the
instrument and disconnect the power cord.
power cord.
page 1-23.
"Lid top cover"
on
226