Table 1-3. Voltage Configuration Plugs
Frequency
Voltage
(Hz)
100 V (± 10%)
48-66
120 V (± 10%)
48-66
230 V (± 10%)
48-66
2. Check the fuses installed on the board. If
they are incorrect for your GC line voltage,
install the correct fuses.
3. Slide the AC board assembly into the chassis
and over the studs. Install the top nut on the
AC board bracket onto its threaded stud.
4. Tighten the remaining nut on the AC board
bracket.
5. Connect the cables and the oven heater
spade lugs.
Released: April 2004
Max. power
Dedicated power
consumption
line requirement
1440 W
15 amp receptacle
1440 W
15 amp receptacle
2000 W
7.5 amp receptacle
6850 Series Gas Chromatograph Service Procedures
Part number
G2630-60740
G2630-60830
G2630-60840
254