Agilent Technologies 6850 Series Service Information page 159

Gc system

Advertisement

5. Loosen the two bottom screws on the detec-
tor electronics cover.
6. Unscrew the top screw on the cover.
7. Remove the detector electronics cover.
Released: April 2004
8. Press down on the orange tabs on the TCD
Remove screw
Detector
electronics cover
Loosen screws
9. Close the lid.
10. Use a T-20 Torx driver to unscrew the gang
11. Disconnect the detector heater/sensor cable
12. Remove the two T-20 Torx screws securing
6850 Series Gas Chromatograph Service Procedures
Board to release the PRT and filament wires.
block fitting from the flow module.
from the harness.
the TCD assembly in place.
Filament wires
PRT wires
159