13. Reinstallation is the reverse of removal.
When done, check for leaks using the Control
Module.
Direct valve-to-column connector kit
Use this kit to connect your 6850 gas or liquid
sampling valve directly onto a column to avoid
unacceptable peak tailing. To use this kit, your
6850 GC must have a purged packed inlet.
Released: April 2004
This kit contains:
Description
Direct-tube assembly
SS bulkhead union, 1/16-inch to
530 µm column
SS bulkhead union, 1/16-inch to
320 µm column
Ferrule, Vespel/graphite, 1/16-inch 0100-1512
Liner, polyamide, for 530 µm
columns
Liner, polyamide, for 320 µm
columns
SS nut, 1/16-inch
1/8-inch to 1/16-inch reducing
union
Liner tool
6850 Series Gas Chromatograph Service Procedures
Part no.
Qty.
n/a
1
0100-1515
1
0100-1527
1
4
0100-1513
1
0100-1514
1
0100-1511
2
0100-1542
1
18900-20850
1
71