Mounting posts
Screw plugs
(may not
be
present)
Figure 1-51. Assembling the valve driver
block
5. For each additional valve installed, install
two more mounting posts and a valve driver
with O-rings in the same manner as the first.
6. When all the drivers have been installed,
screw on the other end plate with the two
hex screws as shown.
Released: April 2004
O-ring
6850 Series Gas Chromatograph Service Procedures
Replacing the bracket and cabling
NOTE: The valve actuators must be installed
before installing the actuators.
1. Unscrew the valve driver bracket from the
right side of the GC using the two captive
screws.
Figure 1-52. Assembling the valve driver