7. Now secure the first nut by adding the
second M10 nut. Tighten it firmly while hold-
ing the first nut stationary using the 17-mm
open-end wrench.
8. Reconnect transformer cables and reassem-
ble the unit in reverse order. Remember to
reconnect the back panel cooling fan.
This completes replacement of the transformer
and mounting pad. Restore power to the unit and
verify its normal operation.