6. Disconnect the µECD vent tube at the rubber
sleeve.
7. Remove the Torx screw holding the gang fit-
ting onto the detector module, and remove
the gang fitting.
8. Remove the one Torx T-20 screw securing
the top µECD detector cover to the detector
pallet and remove the cover.
Released: April 2004
6850 Series Gas Chromatograph Service Procedures
164